The Reference 3000AE™ with Auxiliary Electrometer, is a high-performance Potentiostat/Galvanostat/ZRA just like the Reference 3000 but the Reference 3000 AE features eight additional electrometers and is the ideal potentiostat for battery, capacitor, and fuel cell development. Key features include: • Eight additional voltage measurements – Ideal for stack testing • Maximum Applied Potential - Up to +/-32V@+/-1.5A or +/-15V@+/-3A • EIS - 10 μHz - 1 MHz
The Reference 3000 AE is a high-current, high-performance potentiostat designed for use by battery, fuel cell and supercap researchers. The additional voltage sensing channels make it especially useful for testing of multiple cells within a stack or measuring auxiliary voltages from devices such as pressure gauges or pH meters.
The Accuracy Contour Plot provides a detailed look at the performance you can expect from your instrument in real-world situations. The results include the cell cable.

Extended Cable Lengths
Extended length cell cables (1.5 and 3 meter) are also available for your convenience.
Auxiliary Electrometer Cables: The standard Auxiliary Electrometer cables are 90 cm. Extended length cables (1.5 and 3 meter) are available.
Below are additional details regarding the capabilities of the Reference 3000 AE potentiostat. Each bullet point contains a list of the type of techniques available for the instrument to run.
Other features
SYSTEM |
|
| Potentiostat | Yes |
| Galvanostat | Yes |
| Zero Resistance Ammeter | Yes |
| Floating (Isolated from Earth Ground) | Yes |
| Cell Connections | 2, 3, 4 5 or 21+ |
| Maximum Current | ±3 A or ±1.5 A @ 32V |
| Current Ranges | 11 (300 pA – 3 A) |
| Current Ranges (including internal gain) |
13 |
| Minimum Current Resolution | 92 aA |
| Maximum Applied Potential | ±32 V |
| Rise Time | <250 ns |
| Minimum Time Base | 3.333 μs |
| Noise and Ripple (typical) | <2 μV rms |
| Weight | 7 kg |
| Dimensions | 20 (W) x 23 (H) x 30 (D) cm |
CONTROL AMPLIFIER |
|
| Compliance Voltage | ±32 V |
| Output Current | >±3 A |
| Speed Settings | 5 |
| Unity Gain Bandwidth | 1100, 330, 50, 5.0, 0.5 kHz |
EIS MEASUREMENT |
|
| EIS | 10 μHz – 1 MHz |
| EIS Accuracy | See Accuracy Contour Plot |
| Voltage AC Amplitude | 3 V maximum |
| Current AC Amplitude | 3 A maximum |
ELECTROMETER |
|
| Input Impedance | >1014 Ω || <0.2 pF |
| Input Current (typical) | <6 pA |
| Bandwidth | >15 MHz at – 3 dB |
| CMRR | >80 dB (100 kHz) >60 dB (1 MHz) |
POTENTIAL |
|
| Applied Accuracy | ±1 mV ±0.2% of setting |
| Applied Resolution | 12.5 μV, 50 μV, 200 μV/bit |
| Measured Accuracy | ±1 mV ±0.2% of reading |
| Measured Resolution | High-Resolution Electrometer: 400 μV, 100 μV, 10 μV, 1 μV/bit High Voltage Electrometer: 1.6 mV, 400 μV, 40 μV, 4 μV/bit |
CURRENT |
|
| Applied/Measured Accuracy | ±5 pA ± 0.05% of range ±0.2% of value (3 A – 3 nA) or 0.5% of value (300 pA) |
| Applied/Measured Resolution | 0.0033% full-scale/bit |
| Bandwidth | >10 MHz (3 A – 3 mA) >0.15 MHz (30 μA) |
Carrera 10 # 21-15 edificio Invertobón - Oficina 406 Pereira, Colombia
310 404 3123