The Interface 1010T™ #992-00126 (992-126) is a perfect potentiostat for use in undergraduate and graduate teaching labs.
Key features include:
1. Maximum Current - ± 100 mA
2. Maximum Applied Potential - ± 5 V
3. EIS - 10 μHz - 20 kHz
The Interface 1010T is an entry-level potentiostat designed for undergraduate and graduate teaching labs. It provides all of the capabilities needed for getting students introduced to physical and analytical electrochemistry. It even provides for potentiostatic EIS up 20 kHz.
See the Specifications tab for the instrument specifications. The Interface 1010T can be upgraded to a fully-featured Interface 1010E with an infield upgrade. This allows you to get a low-cost potentiostat for teaching purposes that can later be upgraded to a fully-capable research-grade potentiostat able to perform experiments for a wide variety of applications.
SYSTEM | |
Potentiostat | Yes |
Galvanostat | Yes |
Zero Resistance Ammeter | Yes |
Floating (Isolated from Earth Ground) | Yes |
Cell Connections | 2, 3, or 4 |
Maximum Current | ± 100 mA |
Current Ranges | 6 (1 uA – 100 mA) |
Current Ranges (including internal gain) | 8 (10 nA – 100 mA) |
Minimum Current Resolution | 0.3 pA |
Maximum Applied Potential | ± 5 V |
Rise Time | <1 μs |
Minimum Time Base | 1 ms |
Noise and Ripple | <20 μV rms |
Weight | 2.4 kg |
Dimensions | 24 (W) x 6 (H) x 27 (D) 27 cm |
CONTROL AMPLIFIER | |
Compliance Voltage | ±20 V (@ 1A) |
Output Current | >±1 A |
Speed Settings | 5 |
Unity Gain Bandwidth | 980, 260, 40, 4, 0.4 kHz |
EIS MEASUREMENT | |
EIS | 10 uHz – 20 kHz |
EIS Accuracy | 99% |
AC Amplitude | 2.33 V maximum 17.8 μV minimum |
ELECTROMETER | |
Input Impedance | >1012 Ω |
Input Current (typical) | <20 pA |
Bandwidth | >10 MHz |
CMRR | >80 dB (10 kHz) >60 dB (1 MHz) |
POTENTIAL | |
Applied Accuracy | ±1 mV ±0.2% of setting |
Applied Resolution | 200 μV, 50 μV, 12.5 μV/bit |
Measured Accuracy | ±1 mV ±0.3% of reading |
Measured Resolution | 1 μV, 10 μV, 100 μV, 400 μV/bit |
CURRENT | |
Applied Accuracy | ±5 pA ±0.3% of range |
Applied Resolution | 0.0033% full-scale/bit |
MEASURED CURRENT | |
Measured Accuracy | ±5 pA ±0.3% of range |
Measured Resolution | 0.0033% full-scale/bit |
Bandwidth | >10 MHz (1A – 100 μA) >1.5 MHz (10 μA) >0.15 MHz (1 μA) |
Below is a table of all materials Gamry Instruments supplies in the Laboratory Course in Electrochemistry to perform all eleven experiments for ten pairs of students.
Part number | Description | Quantity | |
820-00005 | Sample of mild steel | 30 | |
930-00015 | Ag|AgCl reference electrode | 1 | |
932-00024 | Platinum working electrode | 1 | |
932-00009 | Platinum microlectrode | 1 | |
935-00065 | Micro stirring bar | 1 | |
935-00123 | Carbon screen-printed electrode | 36 | |
935-00124 | Platinum screen-printed electrode | 60 | |
972-00065 | Cuvette cap | 1 | |
987-00099 | DigiElch student edition | 1 | |
988-00049 | Student edition book | 20 | |
988-00050 | Teacher edition book | 1 | |
990-00193 | Dr. Bob’s Cell kit | 1 | |
990-00195 | Polishing kit | 1 | |
990-00196 | EuroCell kit | 1 | |
990-00419 | AC Dummy Cell | 1 | |
990-00420 | Screen-printed electrode cell stand for 4 mm banana plugs | 1 | |
990-00421 | Screen-printed electrode adapter for 4 mm cable | 1 |